Multifunctional: This root canal preparation machine is two - in one for root canal preparation and root canal length measurement.
Enhanced Experience: The 800mAh battery allows for longer treatment sessions, while the 16:1 contra-angle head provides a wider field of view.

Indication Function: it should be able to display the battery power, and should prompt when the battery power is low; it should be able to indicate the speed, torque and motor steering.
Identification Of Apical Foramen: When the device detects apical foramen, it should be able to give visual an auditory prompt.

Positioning Accuracy: When using the device to locate the root apical foramerthe difference between the positioning position and the actual position should be within the range of ±0.5mm.
Smart Adjustable: The preparation machine shall have an operating control device with adjustable speed. the torque is 0.3-4.0Ncm adjustable.

Program Selection: The root canal preparation machine has 10 working modes for users to save the two, parameters of torque and speed.
Silent Operation During Treatment: The A-weighted sound pressure value produced by the root canal endo motor should not exceed 60dB.

Parameter:
| Speed Range |
140-550r/min |
| Torque Range |
0.3-4.0N.cm |
| Preset Programs |
9 Mermory Modes |
| Working Modes |
4 Modes |
| Standard Handpiece |
16:1 |
| Weight |
137g |
| Power |
AC 110-220V, 50-60Hz |
| Output Voltage |
DC 5V |
| Battery Capacity |
800 mAh |
Packing List:
|
Item
|
Pcs
|
|
Host
|
1pcs
|
|
File Clip
|
1pcs
|
|
Silicone Sleeve
|
2pcs
|
|
Bending Machine Head
|
1pcs
|
|
Lip Hook
|
2pcs
|
|
Adapter
|
1pcs
|
|
Measuring Line
|
1pcs
|
|
Silicone Tray
|
1pcs
|
|
Instruction Manual
|
1pcs
|
|
Certificate
|
1pcs
|
Installation:
Note: During installation or disassembly of the machine, please keep the machine in a non-working state.
1. Connection of charging base
One end of the hand holds the charging base, and the other end holds the adapter connection port, and the aligns the base with the corresponding charging socket and inserts the adapter, The adapter connection port is fully inserted into the charging socket of the base, and no interface is exposed, which means it is connected in place. if there is no need to charge, you can directly pull out the adapter
2. Installation and disassembly of contra angle handpiece
When connecting the contra angle to the host, align the positioning sheath of the contra angle with the positioning slot of the host and insert it until you hear a "click". When removing the contra angle, pull it out in a straight direction.
3. Installation and removal of files
Insert the endo file into the rotating shaft hole of the contra angle, rotate it slightly, and find the position corresponding to the semicircular part of the file. After finding the position, insert a part first, do not pres the button on the contra angle, and then use a slightly larger one. Forcefully press the file into the contra angle head, and when you hear a “click”, the file is connected to the contra angle. To remove the file, press and hold the button on the top of the brake while puling the file out of the handpiece.
When installing the file, be sure to pull and press the file a few times back and forth to confirm that the file is installed before using it. Please keep the file clean, the file vibrate when dirt enters the contra angle Requirements for the selection of root canal files: root canal files(customer-provided), you should choose root canal files with registration certificates on the market and also meet the following requirements: root canal file diameter2.33~2.34mm.
Trouble Shooting:
|
Fault state
|
Possible reason
|
Approach
|
|
Continuous beeping sound after starting the motor handle
|
Start the motor handle to emit a continuous beep to indicate that the motor handle is in reverse mode
|
Stop the motor handle and set the working mode to forward mode
|
|
Contra-angle handpiece calibration failed
|
Calibration failure due to excessive resistance of the contra-angle handpiece
|
clean the contra-angle handpiece, fill it with oil and then re-calibrate it.
|
|
Motor handle gets hot
|
Excessive time in reciprocating mode
|
Stop using it, and start using it after the temperature of the motor handle
|
|
Shorter battery life after charging
|
The battery capacity becomes smaller
|
contact your local dealer or our company
|
|
no sound
|
The volume is set to Vol,0 mute state
|
The volume is set to Vol.1, vol.2 or Vol.3 as required
|
|
The continuously rotating file gets stuck in the root canal
|
incorrect parameter setting the load torque of the file is too high
|
Set the working mode, select reverse mode, start the motor and take out the file carefully
|