18L/23L Class B Dental Autoclave Steam Sterilizer, Fully Intelligent, With Built-in Printer, LCD Display, 121℃/134℃

Istar Dental SupplySKU: MT-AC-03-1

Volume: 18L
Price:
Sale price$1,989.00
Chat on WhatsApp
Free Shipping
United States: in 5-15 days. 0ther Countries: in 15-20 days.
Return & Refund
Returnable within 30 days for return or refund.
Free Registration
Sign up and get 5% off your first order.
Warranty Time
The warranty period for our dental equipment is 6 months

Description


Comply with European B-class sterilisation standard, wide range of application

Comply with EN 13060 safety standard, with three times pulsation vacuum drying function, the vacuum degree can reach -0.9bar (-0.09Mpa). The equipment is capable of eliminating residual moisture on instruments to <0.2%, and is suitable for sterilisation of a wide range of instrument types, including packaged, unpackaged, solid, Class A hollow, Class B hollow, porous structures and instruments with internal piping.

Intelligent control system

Adopting 16-bit imported microprocessor, it provides full computer control operation, which makes the equipment run stably and smoothly. This intelligent system is suitable for efficient disinfection and sterilisation of all kinds of instruments in stomatology, ophthalmology, operating theatres and laboratories. Equipped with B&D test and vacuum test procedure to detect steam penetration and ensure the highest standard of sterilisation.

Real-time monitoring and fault detection function

The clear digital LED display shows the real-time status of the equipment and has a perfect fault detection system. Users can grasp the working status of the equipment at any time, the system will automatically alarm when encountering abnormal conditions, to help users quickly troubleshoot the problem, to ensure stable operation of the equipment.

Multiple safety protection design

Autoclave dental is equipped with double door lock linkage protection, the door can't be opened when there is pressure in the chamber, avoiding the safety hidden danger caused by misoperation. This design adds more security to the process of use, especially suitable for demanding medical environments.

Complete disinfection record management

The built-in printer supports automatic printing of disinfection process data and date records, and the optional USB interface facilitates the export of data. This feature allows the user to establish and manage complete disinfection records to meet the high standards of medical and laboratory management requirements.

High-performance pot and automatic cleaning function

The pot body is made of 2.5mm thick stainless steel by one-time stretch moulding, together with the automatic cleaning device of jet steam generator, which makes the temperature and pressure distribution inside the pot more uniform and ensures that the sterilisation process is thorough and effective. The automatic cleaning function reduces the frequency of manual maintenance and ensures long-lasting and stable equipment.

Waste water management and alarm system

Built-in wastewater tank water full alarm system to avoid recycling of wastewater, to ensure that each disinfection process is clean and hygienic, to improve the sterilisation effect, to avoid the risk of secondary pollution caused by wastewater.

 

Parameter:

Rated voltage

AC220V/50Hz

Rated power

2000W

Door closing method

Manual

Sterilization cycle classification

Class B

Sterilization temperature/pressure

121℃/100120KPa (under standard atmospheric pressure)

134℃/200220KPa (under standard atmospheric pressure)

Volume

18L/23L

Inner cavity size

φ249*355mm/φ249mm*450mm

Sterilization chamber configuration

5 layers, 3 trays

Water supply system

Clean water tank volume :3L

Waste water tank volume :3L

Sterilization record

With built-in printer

Appearance size

640*445*395mm

Packaging size

740*550*500mm

Net weight

45kg

Gross weight

53kg

Special sterilization

Inactivation of HIV, HBV, mad cow virus and spores

Drying procedure

Strong vacuum drying, residual humidity of the instrument <0.2%

Testing procedure

Bowie & Dick steam penetration test, helix spiral test, vacuum test

Safety device

Safety valve, pressure lock, fault self-checking system

Cleaning procedure

Automatic cleaning of internal pipes and steam generator (automatic cleaning every 30 times)

Precautions:

1. Before adding water, be sure to turn on the power supply of the steriliser, and add water not more than 2000mL.
2. If the water shortage indicator lights up and alarms during the sterilisation process, because the water in the storage tank can still satisfy one working cycle, please add water after the sterilisation is completed.
3. Sterilised instruments should be placed on the instrument tray, and there should be a gap between the instruments to facilitate the circulation of steam. Use the supplied instrument tray holder to transport instruments into the sterilisation chamber to prevent burns.
4. If the sterilisation room is hot and there is still water vapour, it is normal that there will be a lot of resistance to closing the door. We recommend opening the door to allow the water vapour to dry out and then closing the door, or pushing the door inwards when closing.
5. Remove instrument trays from the sterilisation chamber using the supplied instrument tray holder to prevent burns!
6. If the sterilised items are removed immediately after the end of the procedure, a small amount of condensation may be found on the sterilised items.
7. After opening the door, the inner chamber of the device, the inner side of the door, the sterilised items, as well as the instrument trays and racks are in a high temperature state, in order to prevent scalding necessary protective measures must be taken to protect the safety of the operator.
8. When the steriliser is no longer in use, switch off the main power switch. (Power switch light off).
9. Please clean the filter in the sterilisation chamber frequently to avoid the filter being clogged and affecting the drainage and exhaust.
10. Do not attempt to open the door of the steriliser when the pressure indicator does not show ‘00’.

Problem And Troubleshooting:

Alarm Code Display

Alarm

Fault Content

Handling Method

E1

"Toot" long sound

Internal temperature sensor error (short circuit)

Check whether the internal temperature sensor is touched

E2

"Toot" long sound

The pressure exceeds 240KPa

Check the pressure sensor and vacuum pump

E3

"Toot" long sound

External temperature sensor error (short circuit)

Check the external temperature sensor

E4

"Toot" long sound

The internal temperature is too high during the pressure boost

Check whether the internal temperature sensor is pressed

E5

"Toot" long sound

Pressure cannot be released

Please clean the pan filter; Ensure smooth exhaust

E6

"Toot" long sound

The door switch pops open while working

Please check that the door handle is turned to the end

E7

"Toot" long sound

The evaporator temperature sensor is incorrect

Check the evaporator temperature sensor

E8

"Toot" long sound

Boost failure

Check the water inlet pipe and air leakage

EH

"Toot" long sound

Evaporator does not heat

Check the evaporator heating rod

EF

"Toot" long sound

The pressure is not higher than 0KPa

Verify that distilled water is being used and check the evaporator

EL

"Toot" long sound

The pressure does not exceed 20KPa

Check the solenoid valve for air leakage or water pump

Eo

"Toot" long sound

The pressure did not exceed 65KPa

Check for air leakage and change the pump without air leakage

E9

"Toot" long sound

Heat preservation and pressure protection failed

Please check for air leakage

EC

"Toot" long sound

quit midway

Cancel Alarm

Payment & Security

Visa Mastercard American Express PayPal Discover

Your payment information is processed securely. We do not store credit card details nor have access to your credit card information.

All of our products come with free shipping, so please don't worry about shipping.

We have put a lot of focus on making sure the items get delivered to our customers as quickly as possible.

The total time is the sum of the processing time and the shipping time. Processing time = the time between when you place an order and when we ship it; Shipping time = the time between when your order has been shipped and deliveto you;

All products have a 1 day processing time

For most products, processing time is 1 day and shipping time is 7-20 days.
In the United States it usually takes 7-15 business days to receive your order; in other countries it can take up to 15-20 days to receive your order.

For large dental equipment, such as dental chairs, the processing time is 1-3 days and shipping time is usually 30-40 days.

All of this information is described inside the product details, if you would like to confirm a more precise time, please contact us and we will calculate it for you as soon as we receive your message.

Any other question, please contact us at info@istardentalsupply.com

Cancellation:

If you wish to cancel your order before shipment, please contact us within 12 hours of placing your order using the contact form or email info@istardentalsupply.com. If you notify us after 12 hours of placing your order, we may not be able to cancel it.

After we receive your message, we will promptly cancel your order and refund the money to your account within 48 hours.

 

Return & Exchange:

1、Return and exchange terms

To qualify for a return, your merchandise must be unused and in the same condition as when you received it.

(1) Your merchandise must be in its original packaging.

(2) You must notify us of your intention to return the merchandise within 15 days of receiving it.

(3) Returned items must be received within 20 days after the return request form is issued.

Please note: For health and hygiene reasons, we do not accept returns of opened dental handpieces, dental intraoral cameras, dental air polishers, and similar items unless defective.

 

2、Defective Items Purchased Online

If you receive an item you believe to be defective, or if there is an issue with a purchase you made, please contact us via the contact form or email. A member of the Istar team will assist in resolving the issue or provide you with the necessary return forms.

 

3、Return and Exchange Information

If you wish to return or exchange items, please provide the following information to us.

(1) Order number

(2) Proof of payment

(3) Product images

 

4、Return/Exchange Process

Contact us via the contact form or email info@istardentalsupply.com, and we'll respond within 12 hours.

 

Return Shipping Fee

Unless otherwise specified, you are responsible for the cost of returning items to us. We recommend all returns be sent back to us via a tracked service, and if the value of the goods exceeds $50, please use an insured service. We are not liable for any returns lost in transit.

Please Note: All returns for both unwanted or faulty items must be logged via our ticketing system. Unauthorised returns unfortunately will not be processed.

 

Refunds

Once we receive your item, we will inspect it and notify you that we have received your returned item. We will immediately notify you on the status of your refund after inspecting the item. If your return is approved, we will initiate a refund to your credit card (or original method of payment).

You will receive the credit within 3-5 workings days depending on your card issuer's policies.

 

After-Sales Contact Information

Tel: +1 430 422 6876

Email: info@istardentalsupply.com

Note: Please contact us before returning any items, otherwise any returns will be refused.

You may also like