Product name: Air Polisher
Intended use: The ISDENT air polisher is intended use for removing tartar, plaque on the surface of teeth and in periodontal pockets, polishing the teeth surface.
Features:
1. Stable Water Supply
The air polisher is designed to maintain a steady flow of water, even at a low input pressure of 0.15 MPa. It provides at least 20 mL of water per minute, ensuring consistent performance throughout procedures.
2. High-Temperature Sterilizable
Built to withstand high-temperature sterilization, the handpiece can endure up to 250 sterilization cycles without any signs of rust or damage. This makes it highly durable and easy to sanitize.
3. Rotating Nozzle
With a rotatable nozzle, this air polisher can easily reach and clean different areas within the mouth, allowing for thorough and efficient polishing from any angle.
4. Ergonomic Design
The air polisher features an ergonomic and attractive design that makes it comfortable to hold and operate, reducing fatigue for the clinician during long procedures.
5. Transparent Sand Jar Lid
The clear lid on the sand jar allows for quick checking of remaining powder levels, so you’ll know exactly when it’s time to refill.
6. Fine Nozzle Design
Designed with a fine nozzle, it offers a broad view of the work area and provides precise control for accurate polishing.
7. Anti-Clogging Design
This air polisher has a smart water and air separation system that prevents clogging, ensuring smooth operation without interruptions.
Parameter
|
Pressure of water supply:
|
0.15MPa to 0.4 MPa
|
Working pressure:
|
0.25MPa to 0.4 MPa
|
Noise:
|
≤80dB
|
Usage:
|
Can be used for basic oral cleaning, plaque control, supragingival and subgingival treatment of implanted
|
Installation
1. Schematic diagram of the device
①nozzle ②handpiece ③device body
④polisher container ⑤cap ⑥ connector
2. Handpiece installation
Connect handpiece by arrow direction (See Figure 2)
3. Filling the powder
Firstly unscrew the cap of container counterclockwise, fill the powder in the container, and then install the cap clockwise(See Figure 3).
4. Subgingival nozzle installation (match the subgingival polisher handpiece)
5. Connection of polisher handpiece and dental scaler
Maintenance
1. There are two waterproof O-rings at the connecting end of the device body and the
handpiece. As they need to be plugged and unplugged repeatedly, once they are damaged or excessively worn, the waterproof O-rings should be replaced in time.
2. There is a sealing O-ring at the connecting end of the device body and container. As the container cap needs to be removed repeatedly, once it is damaged or excessively worn, the sealing O-ring should be replaced in time.
3. After each use, please pour out the remaining powder in the container, and use compressed air or a clean, dry cloth to clean the outer surface of the handpiece, container or threads of container cap, O-ring and the residual powder in the container, and finally use Air Polisher for 5 to 8 seconds without powder to clean the internal tube.
Accessories list
No.
|
Name
|
PCS
|
1
|
Air Polisher
|
1
|
2
|
AH-3 handpiece
|
1
|
3
|
Dental quick connector
|
1
|
4
|
O-ring 6*1.5 mm
|
2
|
5
|
O--ring 11*1.5 mm
|
2
|
6
|
O-ring 33*1.5 mm
|
2
|
Troubleshooting
Error
|
Possible causes
|
Solutions
|
Air leakage of container
|
O-ring is aging
|
Replace the O-ring
|
Liquid leakage of handpiece
|
O-ring is aging
|
Replace the O-ring
|
Air or liquid leakage of handpiece
|
O-ring is aging
|
Replace the O-ring
|
When operating, powder, air or water do not escape from the polisher nozzle
|
There are no air or water
|
Check if there is air or water in dental quick connector
|
Handpiece is blocked by powder
|
Pull out the handpiece
and check if air or powder is sprayed out from the device,if powder, air and water can be sprayed out, please use a steel wire to dredge the handpiece, and then clean the handpiece with compressed air
|
No powder, with air supply
|
There is too little powder in container
|
Check if there is too little powder in the container
|
The supply of air is wet and the powder is caking in container
|
Empty and dry the container, refill the powder,and ensure a dry and clean air source
|